Abstract

This work reports the development of a sequential injection/stripping analysis method for the determination of trace Pb(II) at injection-moulded electrochemical fluidic cells. Conducting carbon fibre–loaded polystyrene electrodes were integrated within the plastic cells. The flow-through cells were incorporated into a home-made sequential injection analysis (SIA) manifold. Different experimental parameters for the detection of Pb(II) were investigated including the type and concentration of the supporting electrolyte, the conditions of the stripping step, the volume of the sample, the flow rate and the accumulation potential. The LOD for Pb(II) was 0.5μgL−1, the within-cell % relative standard deviation (n=8) was 3.1% and the between-cell % relative standard deviation (n=5) was 8.9% for 25μgL−1 Pb(II). The cells were applied to the determination of Pb(II) in tapwater and a phosphate fertilizer sample.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.