Abstract

The recoil fragments from an alpha-emitting source may cause significant detector contamination. The simplest method to avoid the problem is to use a source coating. The influence of Mylar film coatings of different thickness on the resolution, efficiency and minimum detection level of an alpha spectrometry system was investigated. No significant deterioration of the system characteristics was observed when using thin films of about 0.25 mg/cm 2.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.