Abstract
A rapid, simple and sensitive method was developed for the determination of para-aminohippuric acid (PAH) in rat plasma using liquid chromatography tandem mass spectrometry (LC-MS-MS). Acetaminophen was used as the internal standard. Chromatographic separation was performed using a Symmetry C 18 column and the mobile phase was composed of A: 2 mM ammonium formate and 0.1% formic acid in water and B: 2 mM ammonium formate and 0.1% formic acid in acetonitrile (ACN) (A:B, 30:70, v/v). Detection was performed on a triple–quadrupole tandem mass spectrometer using positive ion mode electrospray ionization (ESI) in the multiple reaction monitoring (MRM) mode. The MS/MS ion transitions monitored were m/ z 195.2 → 120.2 and 152.1 → 110.1 for PAH and acetaminophen, respectively. Good linearity is observed over the concentration range of 0.1–500 μg/ml. The method was proved to be accurate and reliable and was applied to a pharmacokinetic study in rat.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.