Abstract
Abstract A gas-solid chromatographic method is presented for the analysis of oxygen, oxygenhelium, and oxygen-carbon dioxide mixtures. The method utilizes a 16’ Porapak Q chromatographic column, hydrogen carrier gas, and a thermal conductivity detection system. It is applicable to the analysis of nitrogen, helium, argon, carbon monoxide, and nitrous oxide. Dry Ice conditions are required for the oxygen analysis; carbon dioxide is determined at ambient temperatures. Details for the assembly of a suitable chromatograph and a procedure for sampling gas cylinders are presented. Samples of medicinal gases were analyzed by the proposed method and by the Orsat absorption method. The results by both methods were in good agreement.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.