Abstract

Organophosphorus flame retardants (OPFRs) are new types of environmental pollutants, therefore the rapid and sensitive detection of OPFRs is a very important objective. A new experimental phenomenon was found in which tris(2-chloroethyl)phosphine (TCEP), a type of OPFR, could effectively enhance the signal of the luminol-H2 O2 chemiluminescence (CL) system. Combined with the controllability of flow injection analysis, a rapid, stable, and sensitive CL method was established. The CL intensity responded linearly to the concentration of TCEP in the range 0.5-100 μg/L (R2 = 0.999) with a low detection limit of 33 ng/L. Relative standard deviation (RSD) was 2.2% (n = 7, c = 100 μg/L). Water samples were labelled and recycled with RSDs of 1.1-5.7% and recoveries of 88.7-116.1%. Based on these results, this study established a new CL detection method for the environmental pollutant TCEP.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.