Abstract

An on-line method for the determination of airborne organophosphate esters based on dynamic sonication-assisted solvent extraction and large-volume injection (LVI) gas chromatography with nitrogen–phosphorous detection is introduced. The LVI is performed with a programmed-temperature vaporizer. The entire extracted fraction of 800 μl (hexane–methyl-tert.-butyl ether, 7:3, v/v) is introduced directly into the GC system without any clean-up step following extraction. The extraction and analysis step were completed in less than 15 min. The limit of detection of the investigated organophosphate esters was established to be in the range of 5–32 pg/filter. The correlation coefficients ( r 2) were investigated in the linear range study of the entire system and established to be ∼0.9900 for all the investigated organophosphates esters. Applications of the method was demonstrated with the extraction of air samples collected onto glass fiber filters from different indoor environments. Six organophosphate esters were found at the levels 0.4–138 ng/m 3.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call