Abstract

Studies were conducted on separation of organic substances with different boiling points and polarity when simultaneously present in water by capillary gas-liquid chromatography on a Kristall 2000 m chromatograph. The effect of the column phase and temperature conditions on the separation factor and component exit time was investigated. Column temperature programming was used to reduce the component, especially heavy component, retention time. The chromatography regime was selected and a fast method was developed for determining components with different volatility in water during one analysis with sufficiently high reproducibility and reliability of the results. It was shown that the time of the analysis decreased by more than two times with the same accuracy as with the standard methods. The method can be recommended for fast analysis of wastewaters for making a decision on the treatment method.