Abstract

A new practical and time-saving ultra-high performance liquid chromatography (UHPLC) method has been developed for determining the organic impurities in the anthraquinone color additives D&C Violet No. 2 and D&C Green No. 6. The impurities determined are p-toluidine, 1-hydroxyanthraquinone, 1,4-dihydroxyanthraquinone, and two subsidiary colors. The newly developed UHPLC method uses a 1.7-μ particle size C-18 column, 0.1 M ammonium acetate and acetonitrile as eluents, and photodiode array detection. For the quantification of the impurities, six-point calibration curves were used with correlation coefficients that ranged from 0.9974 to 0.9998. Recoveries of impurities ranged from 99 to 104%. Relative standard deviations ranged from 0.81 to 4.29%. The limits of detection for the impurities ranged from 0.0067% to 0.216%. Samples from sixteen batches of each color additive were analyzed, and the results favorably compared with the results obtained by gravity-elution column chromatography, thin-layer chromatography, and isooctane extraction. Unlike with those other methods, use of the UHPLC method permits all of the impurities to be determined in a single analysis, while also reducing the amount of organic waste and saving time and labor. The method is expected to be implemented by the U.S. Food and Drug Administration for analysis of color additive samples submitted for batch certification.