Abstract
This paper aims to introduce an algorithm called “sensitivity-based beam number selection (SBBNS)” for fully automated and case-specific determination of an optimal number of equispaced beams in intensity-modulated radiotherapy (IMRT). We tested the algorithm in five head and neck cases of varying complexity. We used direct machine parameter optimization method coupled with Auto Plan feature available in Pinnacle TPS (Version 9.10.0) for optimization. The Pearson correlation test shows a correlation of 0.88 between predicted and actual optimal number of beams, which indicates that SBBNS method is capable of predicting optimal number of beams for head and neck cases with reasonable accuracy. The major advantage of the algorithm is that it intrinsically takes into account various case- and machine-specific factors for the determination of optimal number. The study demonstrates that the algorithm can be effectively applied to IMRT scenarios to determine case specific and optimal number of beams for head and neck cases.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.