Abstract

The adsorptive and electrochemical behavior of norfloxacin on a glassy carbon electrode were investigated by cyclic and square-wave voltammetry. Cyclic voltammetric studies indicated that the process was irreversible and fundamentally controlled by adsorption. To obtain a good sensitivity, the solution conditions and instrumental parameters were studied using square-wave voltammetry. In acetate buffer of pH 5.0, norfloxacin gave a sensitive adsorptive oxidative peak at 0.920 V (versus Ag-AgCl). Applicability to measurement of norfloxacin at submicromolar levels in urine samples was illustrated. The peak current was linear with the norfloxacin concentration in the range 5-50 microg ml(-1) urine. The detection limit was 1.1 microg ml(-1) urine.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.