Abstract

In this paper, CE coupled with electrochemiluminesence (ECL) detection using a 76-μm Pt disk as working electrode was developed for nicotine (NIC) determination. The major metabolite of NIC is cotinine (COT), which has a similar tertiary amine structure to NIC. However, there is a carbonyl group attached in the structure of COT, which leads to the great decrease in ECL response. In order to improve the ECL response of COT, NaBH(4) was used for carbonyl reduction. After reduction, NIC and COT were separated and detected by CE-ECL. ECL response plotted with NIC concentration was linear between 5.0×10(-7) and 5.0×10(-5) mol/L (81-8100 μg/L), with LOD of 5.0×10(-8) mol/L (8.1 μg/L). The developed CE-ECL method was applied for NIC determination in urine and cigarette samples.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.