Abstract

Metal organic framework based mixed matrix membranes (MOF-MMMs) were synthesized and applied for dispersive membrane extraction (DME) of four neonicotinoid insecticides (nitenpyram, thiacloprid, imidacloprid, and acetamiprid) in environmental water, combined with high performance liquid chromatography (HPLC) for determination. Several experimental conditions were optimized in detail, involving dosage percentage of MOF, extraction time, sample pH, salinity, type and volume of eluent, and elution time. High sensitivity with limits of detection and quantification were achieved as 0.013-0.064 μg L-1 and 0.038-0.190 μg L-1, respectively, and good precision with relative standard deviations were obtained as 3.07-12.78%. The proposed method has been successfully applied to determine four neonicotinoid insecticides in tap water, surface water, and seawater, satisfactory recoveries of spiked water samples were between 72.50 and 117.98%. Additionally, the MOF-MMMs showed good reusability with the extraction efficiencies almost remaining stable after 14 cycles. The MOF-MMMs based DME followed by the HPLC method can be a promising utility for the determination of neonicotinoid insecticides in environmental water samples, with high sensitivity and convenient operation.