Abstract
The determination of neomycin sulfate by liquid chromatography using a column packed with a poly(styrene-divinylbenzene) co-polymer and pulsed electrochemical detection on a gold electrode is described. The mobile phase consisted of an aqueous solution containing 70 g/l of sodium sulfate, 1.4 g/l of sodium 1-octanesulfonate and 50 ml/l of a 0.2 M phosphate buffer (pH 3.0). Sodium hydroxide was added post-column. The influence of the different chromatographic parameters on the separation was investigated. The method shows good linearity and repeatability and is stability indicating. A number of commercial samples was analyzed using this method and the results were compared with results obtained with the European Pharmacopoeia method and a previously described thin-layer chromatographic method.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.