Abstract

An analytical method for the determination of total N-acetylcysteine in human plasma has been developed, validated and applied to the analysis of samples from a phase I clinical trial. The analytical method consists of plasma digestion with dithiothreitol in order to reduce all the oxidized forms of N-acetylcysteine, and extraction with ethyl acetate followed by determination of levels by an LC–MS–MS method. The intra- and inter-assay precision and accuracy of this technique were good and the limit of quantitation was 50 ng/ml of plasma. The concentration working range was established between 50 ng/ml and 1000 ng/ml. This method has been used in the analysis of approximately 800 human plasma samples from a clinical study with 24 volunteers; the precision of the quality controls was in the range 8.7 to 13.4% and the accuracy was in the range −5.9 to 8.5%, expressed as the RSD and the relative error, respectively.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.