Abstract
A simple, sensitive and specific method for the quantification of N-acetyl-S-(N-methylcarbamoyl)cysteine (AMCC) in urines of the general population and of exposed workers has been developed. Samples were first diluted with phosphate buffer followed by purification by solid-phase extraction (SPE) with SAX columns prior to analysis by liquid chromatography/mass spectrometry (LC/MS). An external standard was used for the quantification together with selected ion monitoring of m/z 219, [M-H]-. The linear calibration curve showed a good correlation (r2 = 0.98, p < 0.001) between 0.1-110 mg/L of AMCC, and the detection limit was calculated to be 2 microg/L. The within-day and between-day precision, calculated for exposed workers and general population AMCC levels in urine samples, were determined as 1.2 and 3.6%, respectively (as relative standard deviation (RSD)).
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.