Abstract

Objective: To establish the method for the determination of N-Acetyl-4, 4'-diaminodiphenylmethane (AcMDA) adduct in the hemoglobin by ultra-performance liquid chromatography tandem mass spectrometry (UPLC-MS/MS) . Methods: The 20 mg hemoglobin sample was weighed into a 15 ml centrifuge tube, adding 20 μg/L internal standard solution AcMDA-D8 10 μl, then hydrolyzed with 1 ml 0.1 mol/L sodium hydroxide solution at 37°C for 0.5 hours, extracted with dichloromethane and concentrated by vacuum concentrator. The residue was dissolved in acetonitrile and detected by UPLC-MS/MS, then quantitative by internal standard method. Results: The linearity of the method was good at the range of 0.05-25.00 μg/L with a correlation coefficient of 0.9996, the detection limit (LOD) and limit of quantitation (LOQ) were 0.07 and 0.2 ng/g Hb, respectively. The recovery rate was ranged from 91.0%-95.4%; the relative standard deviation (RSD) of intra-and inter-batch precision were 4.5%-6.3% and 3.7%-4.4%, respectively. Conclusion: The determination method meet the requirement of GBZ/T 210.5-2008.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.