Abstract

To implement and validate an analytical method by ultra-performance liquid chromatography-tandem mass spectrometry (UPLC MS/MS) to quantify mycophenolic acid (MPA) in kidney transplant patients. Quantification of MPA was performed in an ACQUITY UPLC H Class system coupled to a Xevo TQD detector and it was extracted from plasma samples by protein precipitation. The chromatographic separation was achieved through an ACQUITY HSS C18 SB column with 0.1% formic acid and acetonitrile (60:40vol/vol) as mobile phase. The pharmacokinetic parameters were calculated by non-compartmental analysis of MPA plasma concentrations from 10 kidney transplant patients. The linear range for MPA quantification was 0.2-30 mg/L with a limit of detection of 0.07 mg/L; the mean extraction recovery was 99.99%. The mean intra- and inter-day variability were 2.98% and 3.4% with a percentage of deviation of 8.4% and 6.6%, respectively. Mean maximal concentration of 10 mg/L at 1.5 h, area under the concentration-time curve of 36.8 mg·h/L, elimination half-life of 3.9 h, clearance of 0.32 L/h/kg and volume of distribution of 1.65 L/kg were obtained from MPA pharmacokinetics profiles. A simple, fast and reliable UPLC-MS/MS method to quantify MPA in plasma was validated and has been applied for pharmacokinetic analysis in kidney transplant patients.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call