Abstract
A new approach for the calibration of SEC-light scattering (SEC-LS) setups is proposed, which requires solely the molar mass of a reference polymer. Neither the specific refractive index increment of the calibrant nor of the analyte is required. Comparison of the molar masses derived in different solvents for a large number of chemically different polymers shows that the new approach yields the same molar masses as if molar masses were derived using dn/dc to calibrate the light scattering setup. The approach therefore allows easier determination of molar masses by SEC-LS.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.