Abstract

Determination of moisture content in lyophilized solids is fundamental to predict quality and stability of freeze-dried products, but conventional methods are time-consuming, invasive and destructive. The aim of this study was to develop and optimize a fast, inexpensive, noninvasive and nondestructive method for determination of moisture content in lyophilized mannitol, based on an NIR micro-spectrometer instead of a conventional NIR spectrometer. Measurements of lyophilized mannitol were performed through the bottom of rotating glass vials by means of a reflectance probe. The root mean standard error of prediction (RMSEP) and the correlation coefficient (R²pred), yielded by the pre-treatments and calibration method proposed, was 0.233% (w/w) and 0.994, respectively.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.