Abstract

Simultaneous multielement determination of minor and trace elements in the urine reference sample by an ICP-MS/AES combined system is described. The biological reference sample of NIST SRM 2670 Freeze-dried Human Urine (elevated) was used for analysis. The sample preparation was kept to a minimum: the urine sample was diluted 25-fold with 0.1 M nitric acid, in which Ge, In and Tl were added as the internal standard elements. Eighteen elements at the μg/ ml-ng/ml levels were simultaneously determined by the present combined system. In addition, the present combined system was also applied to the correction of polyatomic interference in the determination of chromium in SRM 2670, where interference with 52Cr due to 40Ar12C was corrected by estimating the correction factor of 40Ar12C from correlation between the ion count of 40Ar12C and the emission intensity of carbon in the sample.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.