Abstract

A significant problem with most field dental units is that their suction is too weak to effectively remove debris from the mouth. The purpose of this study was to determine the minimum clinically acceptable suction level for routine dentistry. A vacuum pump was connected to a high-volume dental evacuation line in a simulated clinical setting and different suction airflow rates were evaluated by nine evaluator dentists for their capability to effectively remove amalgam debris and water. Airflow levels were rated as "clinically acceptable" or "clinically unacceptable" by each evaluator. Data were analyzed using a chi2 test for trend. Analysis indicated a significant linear trend between airflow and ratings (p < 0.0001). The first airflow level considered by all evaluators as producing clinically acceptable suction was 4.5 standard cubic feet per minute (0.127 standard cubic meters per minute). This value should be the minimum level required for all military field dental units.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.