Abstract
The photometric determination of manganese described is based on the oxidation of the Mn to permanganate by means of persulphate, reaction with tetraphenylarsonium chloride, extraction of the complex with dichloroethane from 0.7 N sulphuric acid solution and measurement of the absorption of the extract at 528 nm. Manganese can be determined in the ppm-range in iron, steel, nickel, copper, aluminium and magnesium. Interferences are caused by larger concentrations of chromium.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.