Abstract
The voltammetric behaviour of methylmercury at Nafion ® coated glassy carbon electrodes is examined. Cyclic voltammograms recorded at coated electrodes show a dramatic increase of peak currents with respect to bare electrodes, this effect being more evident in perchlorate supporting electrolytes rather than in chloride containing media. Experimental evidence indicates that ion-exchange preconcentration of the cation CH3Hg is the prevailing process, even if the incorporation of neutral species such as CH3HgCl by hydrophobic interactions cannot be neglected. The application and optimization of pulsed techniques like differential pulse voltammetry and Osteryoung’s square wave voltammetry, indicates the possibility to use Nafion ® coated electrodes for the determination of trace methylmercury at submicromolar levels. The analytical performances can be further improved by using, at the Nafion ® coated electrode, a multiple pulse technique named multiple square wave voltammetry. The application and optimization of this technique to the system under investigation is presented and discussed. © 1999 Elsevier Science S.A. All rights reserved.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.