Abstract
Polycarbonate foils with thicknesses from 0.08 to 2.5 mm containing Ti, Cr, Fe and Ba were analysed by energy-dispersive x-ray fluorescence (EDXRF) spectrometry. Neutron activation analysis (NAA) was used to validate the EDXRF results. The element concentrations ranged for Ti from 0.04 to 0.75%, for Cr from 0.03 to 0.3% and for Fe from 0.25 to 1.30%. The concentration of Ba was about 1%. The results obtained by EDXRF were of the same quality as those provided by NAA, but the NAA method was found to be more accurate for polycarbonate samples containing both Ti and Ba. Copyright © 2000 John Wiley & Sons, Ltd.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.