Abstract
To avoid systematic errors in the sampling and determination of mercury in seawater an integrated sampling-analyzing system has been developed. A 500 ml PTFE-bottle attached to a sampling gear is used as sampling vessel for the cold vapour atomic absorption Spectrophotometry, using nitrogen aeration, SnCl2 reduction, and preenrichment/ purification by amalgamation on finely dispersed gold. The detection limit of this procedure was found to be 0.5 ng·l−1 and the coefficient of variation 4% for 1.5 ng·l−1.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.