Abstract
In this work, a device based on diffusive gradients in thin films (DGT) was evaluated for the determination of Hg(II) in river water. The DGT device was assembled with a cellulose phosphate ion exchange membrane (P81 Whatman) as a binding phase and agarose gel 1.5% (m/v) as a diffusive layer. Laboratory deployments showed that the binding of Hg2+ ([HgDGT]/[Hgsolution]) by P81 membrane was more effective (97%) than the Chelex 100 resin (80%).The effect of ionic strength, pH and potential interfering ions on Hg binding with DGT׳s was investigated. The results showed no significant effect on the binding of Hg(II) at pH range from 3.5 to 8.5 and at an ionic strength range from 0.0005 to 0.1molL−1. Uptakes of 50µgL−1 Hg(II) by P81 membrane were not affected by Fe, Mn, Zn, Cu, Ca and Mg at the concentration range of 200–1800µgL−1. Finally, the DGT device using the P81 as the binding layer was applied for in situ measurements of Hg in river water. For in situ measurements, the labile Hg concentration (from <2 to 13ngL−1) was lower than 10% of the dissolved fraction (from 155 to 446ngL−1).
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.