Abstract
A simple and rapid method has been developed for the determination of total Hg in drainage sediment. The method is performed by using a sequential injection (SI) system with cold vapor generation atomic fluorescence spectrometry (CVAFS) after a closed microwave assisted digestion. Digestion was carried out at a temperature of 140°C for 5 minutes using a mixture of 10% HCl + 50% HNO3 +40% H2O as digestion acid media. Stability experiment revealed that the extracted samples were stable in 24 hours. The SI–CVAFS system was well optimized. At optimized conditions, the detection limit of the method was 0.5 ng/g−1. Relative standard deviation (RSD) for the complete procedure by the analysis of the 10 digestions was 3.7%. The method was validated against 3 drainage sediment certified reference materials: GSD-2, GSD-9, and GSD-10 and successfully applied to the determination of Hg in the sediment of Pi River, China.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.