The analysis of melarsoprol in whole blood, plasma, urine and cerebrospinal fluid is described. Extraction was made with a mixture of chloroform and acetonitrile followed by back-extraction into phosphoric acid. A reversed-phase liquid chromatography system with ultraviolet detection was used. The relative standard deviation was 1% at concentrations around 10 mumol/l and 3-6% at the lower limit of determination (9 nmol/l in plasma, 93 nmol/l in whole blood, 45 nmol/l in urine and 10 nmol/l in cerebrospinal fluid). Melarsoprol is not a stable compound and samples to be stored for longer periods of time should be kept at -70 degrees C. Plasma samples can be stored at -20 degrees C for up to 2 months. Chromatography showed that melarsoprol contains two components. Using nuclear magnetic resonance spectroscopy the two components were shown to be diastereomers which slowly equilibrate by inversion of the configuration at the As atom.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Journal of chromatography. B, Biomedical sciences and applications
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.