Abstract

A novel system for mercury speciation analysis using high-pressure ion chromatography (IC) has been developed and validated. Its chemistry permits separation of the two most abundant forms of Hg in natural waters, soils, sediments, and biota-monomethyl Hg (CH3Hg+) and mercuric Hg (Hg2+)-on the basis of the difference in charge of their respective thiourea (S=C(NH2)20) complexes. Once separated, both species are converted to Hg0 on-line and quantified by cold-vapor atomic fluorescence spectrometry (CVAFS). A column containing thiol-functionalized silica resin installed in the sample loop of the IC system traps Hg2+ and CH3Hg+ from prepared sample solutions without retaining interfering sample matrix components. The resulting matrix-independent chemistry permits external calibration of the system and a high sample throughput ( approximately 6 samples per hour). The system's accuracy has been validated with environmentally relevant reference materials. Figures of merit for the system, an average precision of approximately 2.5% and an absolute detection limit of <1 pg, are comparable to state-of-the-art gas chromatography approaches for Hg speciation analysis (ethylation/GC-CVAFS) and meet or exceed those of all extant LC systems. Low relative detection limits are attainable through preconcentration onto the thiol resin.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.