Abstract

We developed a procedure for the high-precision determination of matrix elements of glasses of the Ga–Ge–As–Se system, including those doped with praseodymium, in the range of concentrations of gallium from 1 to 5 at %, germanium from 16 to 24 at %, arsenic from 14 to 18 at %, and selenium from 57 to 65 at %. A procedure for determining 0.05–0.5 wt % of praseodymium in these glasses by inductively coupled plasma–atomic emission spectrometry is also proposed. The expanded uncertainty (P = 0.95) of the results in determining the matrix elements ranged from 0.05 to 0.1 at %, and for praseodymium, it was from 0.002 from 0.02 wt %. A method for preparing standard solutions necessary to achieve the stated level of uncertainty of the analysis results is described. The minimum sample weight for determining matrix elements is approximately 1 mg, and for the determination of praseodymium, it is approximately 10 mg, which enables analyzing not only bulk glass samples but also fibers made of them.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.