Abstract

Current proposed method allows for the determination of manganese in serum sample using aqueous standard calibration technique on Graphite Furnace atomic absorption spectrophotometer (GFAAS) using deuterium background correction. This method involves determination of manganese from digested serum samples without the use of matrix modifier. Pyrolysis and atomization temperatures are 1200°C and 2200°C respectively. The limit of detection (LoD) and limit of quantitation (LoQ) of the method are 0.0097ng/ml (0.18nmol/l) and 0.032ng/ml (0.58nmol/l) respectively. Validation of the method was carried out using seronorm trace element level-1 serum standard with excellent agreement between measured value and certified value. Accuracy was demonstrated by the spike and recovery study with analytical recovery between 98.8 and 100.6% in serum. The serum reference value for manganese in adolescent girls of rural Konkan region of India range from 4.7 to 215nmol/l.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call