Abstract

A method for the direct determination of manganese in river and sea-water reference materials using electrothermal atomization atomic absorption spectrometry with a tungsten atomizer is described. Interferences caused by the sea-water matrix are investigated and it is shown that these can be eliminated by the addition of ascorbic acid. The method is applied to the determination of manganese in sea-water reference materials and results within one standard deviation of certified values are obtained. The detection limit (3σ) in sea-water is 1.2 pg, corresponding to 0.12 µg l–1 for a 10 µl sample.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.