Abstract

Antibiotics are contaminants at low concentrations and found in water with the range of ng/L. Determination of antibiotics at trace concentrations requires modern and high accuracy equipment. Liquid chromatography mass tandem spectrometry (LC-MS/MS) able to determine contaminants at ppt and ppb levels, which is suitable for quantification of the macrolide and tetracycline group in water. The Analytical method using Agilent Triple Quard 6460 LC-MS/MS system with positive ion mode. Agilent XDB C18 column (1.8 µm × 2.1 × 100 mm) and Agilent Eclipse SDB C18 pre-column (2.1 × 5 mm × 1.8 µm) were used. Mobile phase was acetonitrile and HCOOH 0.1%. Samples were cleaned with Oasis PRiME HLB 3cc SPE column (150 mg). The method was evaluated based on specificity, recovery, repeatability and estimation of the uncertainty. The method detection limit (MDL) for the analytes is 0.03 µg/L. Recovery ranges from 85.23 -117.70%; repeatability (RSDr) is between 2.20% and 12.45%.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.