Abstract

A thermodilution method for measuring distal femoral vein flow in man is presented. The catheter is introduced percutaneously in central direction into the popliteal vein. Tests for accuracy in vitro (r = 0.999, range of true flow 63–2600 ml/min) and in dog experiments (r = 0.994, range of electromagnetic flow measurements 110–780 ml/min) are presented and discussed in relation to the anatomical (dissections on corpses) and functional conditions in the popliteal/distal femoral vein of man. A group of normal persons [7] and of patients suffering from ischaemic lower limb disease [17] were studied at rest and during exercise on an ergometer in supine and sitting positions. In all but two, reliable measurements were obtained. In the normals, a mean resting flow of 114 ml/min (range 82–140 ml/min) was found. The individual coefficient of correlation between load and flow measurements in the normals varied from 0.80 to 0.94 during stepwise increasing load exercise. No serious complications occurred. The method...

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.