Abstract

A gas chromatographic method was developed for the determination of low-molecular-weight carboxylic acids in aqueous samples based on a derivatization procedure compatible with aqueous solutions. The technique uses nitrogen-selective detection with a thermionic-specific detector after derivatization of carboxylic acids as 2- nitrophenylhydrazides. The hydrazides were extracted with ethyl acetate prior to injection into the gas chromatograph. The derivatives appear to be stable in ethyl acetate at 0–5° C for long periods and, therefore, can be stored for analysis at a later date. The detection limits of different short-chain acids are in the range 0.8–1.4 pmol per injected sample. The relative standard deviation is less than 10% at the 1 μM level. Examples of the use of the method are given for the determination of carboxylic acids in anoxic marine sediment pore waters, coastal sea water and Black Sea water samples.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.