Abstract

A method of determining low concentrations of cadmium in marine and fresh water sediments utilizing a reduced PdCl2 chemical modifier for graphite furnace atomic absorption spectrometry (GFAAS) is presented. National Research Council of Canada marine sediment reference materials BCSS 1, MESS 1 and PACS 1 along with National Institute of Standards and Technology Standard Reference Materials 1645 and 1646 were used to determine the precision and accuracy of the method using various palladium-modifier concentrations. The modifier concentration must be approximately 4000 mg l–1 to be effective. The chemical modifier GFAAS method was compared with inductively coupled plasma atomic emission spectrometry (ICP-AES) in the analysis of 199 coastal marine sediments. No evidence of a systematic difference between the two data sets was found at the 95% confidence level. The mean 1σ scatter in the method comparison, from 0.8 to 5.0 µg g–1, is ± 0.27 µg g–1. The ICP-AES limit of detection is degraded from 0.5 to 0.8 µg g–1 by the uncertainty in the sediment inter-element corrections, implying a mean inter-element correction uncertainty of ± 2.5%.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.