Abstract
An acetonitrile–salt stacking method was established for the assay of lipoic acid (LA) in biological samples. Samples were deproteinized with acetonitrile at a final concentration of 60 % (v/v) and then injected hydrodynamically at 3.45 × 103 Pa for 180.0 s. The optimum background electrolyte was found to be 90.0 mmol L−1 pH 9.1 borate buffer. LA could be detected within 35 min at +7.0 kV with satisfactory repeatability (relative standard deviations, RSDs, of migration times and peak areas were both below 10 % for intraday and interday; n = 6/9) and a relatively low limit of detection of ca. 0.5 μmol L−1.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.