Abstract

A carbon fiber microelectrode was used for the electroanalytical determination of Linuron (LIN) in soil extracts. The microelectrode was subjected to an electrochemical pretreatment in order to improve the herbicide adsorption on the electrode surface. With this preconcentration step, detection limits of 80 ng ml(-1) and determination limits of 260 ng ml(-1) were reached. Optimal conditions with respect to accumulation time and potential, scan rate and pH were established. The LIN was determined in a soil sample with the method proposed and the results found were comparable to those obtained by HPLC.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.