Abstract
This study compares the efficiency of blood lead level analysis by graphite furnace atomic absorption spectrometry (GFAAS) and the portable LeadCare Blood Lead Testing System (LCS). Recoveries of two added lead concentrations of 22 and 42 micrograms/dL ranged from 102.4 to 105.5% for LCS and from 96.3 to 97.2% for GFAAS. Measurement of a certified sample (Certified Danish Whole Blood) at a blood lead concentration of 26.2 micrograms/dL gave within- and between-run coefficients of variation which were both approximately 8% by LCS and 2% by GFAAS. Comparison of the tested method (LCS) versus GFAAS from analysis of 76 samples of blood lead collected from workers in different industrial sectors showed imperfect overall correlation (r = 0.95). The LCS is quite suitable for screening purposes, but requires the use of non-frozen blood collected less than 24 h before. Conservative threshold values should be applied when using the LCS for initial screening in the field.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.