Abstract

A simple, selective and sensitive method was developed based on electrothermal atomic absorption spectrometry using Zeeman correction for quantitation of lead, in 100 ml of whole blood sample, as biological indicator for occupational exposure. Confidence parameters and stability of samples were considered. Ashing and atomization temperatures, considered critical, were 700 °C and 1,700 °C, respectively. The levels found during the validation process showed good sensitivity linearity, recovery, precision and accuracy. The stability results presented levels remaining constant for a 15 months period. The variations were not higher than 15% when comparing concentrations in zero time to those obtained after storage period.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.