Abstract

Atomisation from a pyrolytic graphite probe, placed in a hot and constant-temperature HGA 70 furnace, was used for the direct determination of lead in diluted whole blood. Substantial reductions in the classical vapour-phase interference effects by up to 2%m/V magnesium chloride and calcium chloride and 1.5%m/V sodium chloride allowed the use of aqueous standard solutions for analytical calibration. Good agreement with national (UK) mean values was obtained for the analysis of quality control blood samples. The analytical precision is equivalent to that with conventional atomisation, but with improved sensitivity.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.