Abstract

A flow-injection on-line preconcentration-electrothermal atomic absorption spectrometric (ETAAS) method coupled with a coprecipitation method has been developed for the determination of lead in seawater. The combination of two preconcentration procedures, coprecipitation with iron(II) hydroxide and solid-phase extraction with a lead-selective resin, Pb-Spec, allowed the determination of lead at the ng kg(-1) level. Lead in 250 g of a sample solution was collected by coprecipitation with 10 mg of iron. The precipitate was dissolved in 25 ml of 1 mol l(-1) nitric acid; then, a 4-ml aliquot of the sample solution was introduced into the flow-injection system to preconcentrate and separate lead from iron on a Pb.Spec microcolumn. The sorbed lead was eluted with a 1.0 x 10(-4) mol l(-1) EDTA solution. The 30-microl portion of the eluate corresponding to the highest analyte concentration zone was injected into a graphite furnace. The overall enhancement factor was about 200 for 250 g of the sample. The average and standard deviation of ten blank values obtained were 1.7 ng and 0.38 ng, respectively. The recovery was 93.7 +/- 5.0% for seawater spiked with 20 ng kg(-1) lead. The proposed method is applicable to the analysis of seawater for lead at slightly higher levels.