Abstract

A study was performed for the determination of lead in rice grains directly by solid sampling high resolution continuum source graphite furnace atomic absorption spectrometry (HR-CS GFAAS) without digesting sample. The effects of sample amount, pyrolysis/atomization temperatures on the determination of lead in rice were investigated and optimized using a certificated rice flour. The lead concentrations were determined applying 800°C of pyrolysis and 1800°C of atomization temperatures without modifier. LOD (N=10, 3σ) and characteristic mass were 2.3μgkg−1 and 8.1pg, respectively. The certified lead value of a rice flour CRM was found in its uncertainity limits. The lead contents of various rice samples obtained from markets in Turkey were between 0.009 and 0.162mgkg−1 which are in acceptable range. The average lead concentrations in various rice grains taken from the same package were significantly different from each other.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.