Abstract

A laboratory-made atomic fluorescence spectrometer with a capacitively heated tungsten spiral atomizer was applied to the determination of lead in waters. The optical system for fluorescence collection and the electronics for registration have been improved, which made it possible to suppress the emission of the hot spiral and to measure the fluorescence directly from the inner volume of the spiral. The detection limit of lead (3.5 × 10–7 g dm–3) is reduced by a factor of 30 compared with measuring the fluorescence above the spiral. The relative standard deviation is 0.05%(n= 10). An acidity of <1% HNO3, a magnesium content of <0.01 g dm–3 and a calcium content of <0.02 g dm–3 do not affect the lead fluorescence signal.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.