A procedure for extraction of the kinetic energy release distribution from the metastable fragment ion peak shape in an ion trap reflectron time-of-flight mass spectrometer is presented. Peak shapes for a set of discrete values of kinetic energy releases are calculated and their contributions to the measured peak shape determined. Information about the temporal evolution of each component is required in order to get the correct kinetic energy release distribution. This is obtained by carrying out the measurements at different times following ion formation. The method is demonstrated for the metastable decay of aniline ions formed by multiphoton ionization.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Similar Papers
More From: European Journal of Mass Spectrometry
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.