Abstract

AbstractCapillary electrophoresis (CE)/electrochemical detection (EC) for the simultaneous determination of hydrazine and isoniazid has been developed. The electrochemical method uses a novel modified electrode dispersed with ultrafine platinum particles on the surface of a 30 μm carbon fiber microelectrode. The unique characteristic of the Pt‐particles modified carbon fiber microelectrode is its excellent stability. The current measurement for hydrazine is more sensitive than that of isoniazid. Selective determination of trace amount of free hydrazine in isoniazid and its formulation can be achieved at applied potential of 0.5 V.