Abstract

Abstract Objective The aim of current study was the development and validation of a reversed-phase high-pressure liquid chromatography (RPHPLC) method for the quantitative determination of two active ingredients, isohydrafural (IHF) and methyluracil (MU) in fixed dose combination ear drops. Methods: An efficient separation of the two compounds was achieved on a Teknokroma C1 5μm (150*4,6mm) column, with a methanol: water 60:40 (V: V) mobile phase, at 300C temperature and 0.6 mL/min flow-rate. The total analysis time was 5.5 min. Result: The verified validation parameters were: linearity, selectivity, specificity, precision (repeatability and reproducibility), robustness, limit of detection and quantification. Conclusions: Good separation and lack of interference from other chromatographic peaks, rapidanalysis times were obtained. The newly developed chromatographic method can be applied for the analysis of fixed dose combination ear drops with isohydrafural and methyluracil.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.