Abstract
The ion recombination loss is one of the most important correction factors for dosimetry using ionization chamber especially for the measurement of high-intensity pulsed radiation beam. The two-voltage method has been used frequently for the correction of ion recombination loss. Several ways to approximate the method are presented. Comparisons were carried out to check the validity of the approximation. Using four ionization chambers of different types and shapes, measurements were made to obtain the optimal value of the ratio for the two voltages.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.