Abstract
An HPLC method is described for the determination of iodide in serum and urine using ion-pair chromatography with coulometric detection. After adding hexadecyltrimethylammonium chloride, the ions pairs formed with the iodide in the sample are extracted using an organic solvent. The solvent is then evaporated and the dry residue obtained is mixed with an appropriate volume of mobile phase so as to concentrate the sample prior to injection into the chromatograph. For a sample of 0.5 ml of serum, the method features a limit of detection (signal-to-noise ratio of 3) of 0.2 microgram l-1, sufficient to be applied in paediatric assays for the diagnosis of both iodide deficiency and excess.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Journal of chromatography. B, Biomedical applications
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.