Abstract

A robust tandem HPLC method coupling size-exclusion (Shodex Asahipak GS-320HQ) and reversed phase (Vydac 218TP54) columns with ultraviolet detection was developed for quantitative determination of interferon-gamma (IFN-gamma) in Chinese hamster ovary cell culture supernatant. The 2D-HPLC system was linked up by a 6-port 2-position low hold-up volume switch valve. Compared to a commercial ELISA kit for IFN-gamma, the coupled column LC approach was able to detect and quantify soluble IFN-gamma, regardless of the glycoprotein's molecular/conformational variability and sample background. Each LC-LC analysis took 90 minutes inclusive of column regeneration. The relative standard deviation of measurements (n = 5) was less than 3%. The limit of detection (LOD) was determined to be 0.35 microg IFN-gamma.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.